1. Images
Side-view (shadowgraph) photos of drops on a flat surface. PNG, JPG or BMP. You can select several.
Sample photo from J.-W. Lee, Mendeley Data, doi:10.17632/t25zyc857h.1, CC BY 4.0.
2. Settings
Drag the two yellow handles to move or tilt the baseline. Angles update at once.
3. Result
| Circle-fit angle (valid only for small, spherical drops) | – |
| Base diameter | – |
| Drop height | – |
| Baseline | – |
| Fit model | – |
Advanced analyses
These analyses are not part of the free tool. Send us your images or videos and SOREC returns the results with a short report. Request an analysis
How it works
The uneven lighting of the background is removed first, so the drop edge can be found at half contrast everywhere in the image. The edge is traced with sub-pixel accuracy, and the needle is ignored.
The baseline is found from the contact points themselves: the corner where the drop outline meets the substrate, or the neck where the drop meets its own reflection on shiny surfaces. It can be tilted and corrected by hand with the yellow handles.
By default, a Young–Laplace profile (the shape of a real drop flattened by gravity) is fitted to the whole outline, and the angle is read where the profile meets the baseline. Edge points right at the contact line, where the image is blurred, are left out. If the profile does not match the outline, the analyzer falls back to a local tangent fit near the contact line and says so.
The analyzer measures the static (sessile) contact angle. θL and θR are the angles on the two sides of the same drop; on a uniform surface with a level camera they should agree within a degree or two, and a larger difference points to an uneven surface, uneven pinning or a tilted baseline. They are not the advancing and receding angles: contact angle hysteresis needs a moving contact line, measured by adding and withdrawing liquid or by tilting the surface.
Accuracy
We compared the analyzer with two established programs on 80 real photographs of water drops from a published dataset (J.-W. Lee, Mendeley Data, doi:10.17632/t25zyc857h.1), which gives the angles measured with SmartDrop, a commercial instrument, and with Cellamount scope PC. All 80 images were analyzed automatically, with no manual correction. The table gives the mean absolute difference of the mean contact angle.
| Drops (SmartDrop angle) | Images | SOREC vs Cellamount | SOREC vs SmartDrop | Cellamount vs SmartDrop |
|---|---|---|---|---|
| Below 90° | 10 | 2.1° | 2.5° | 0.7° |
| 90 to 150° | 40 | 4.0° | 5.0° | 2.6° |
| 150 to 170° | 23 | 2.5° | 5.6° | 3.5° |
| Above 170° | 7 | 5.6° | 14.1° | 16.6° |
| All | 80 | 3.5° | 5.7° | 3.9° |
Overall, our analyzer differs from Cellamount by slightly less than the two established programs differ from each other, and 63 of the 80 images agree with Cellamount within 5°. It reads on average about 3° lower than SmartDrop between 90 and 170°. A few images with an unclear substrate line still need the baseline corrected by hand. On synthetic images with known angles the mean error is 0.8° for spherical drops, 1.5° for gravity-flattened drops and 2.2° for small, noisy drops.
Tips for good images
Use a backlit side view with the camera level with the surface, the drop in sharp focus and the whole drop in the frame. Make the drop at least 200 pixels wide in the image, but do not let it fill the frame, and avoid strong reflections on the drop edge. The baseline matters most: moving it by one pixel can change a high angle by about a degree, so check that the yellow line touches the contact points. Take particular care with small drops (under about 300 pixels wide) above 120° on non-reflective surfaces, where the automatic baseline tends to sit slightly low and the angle reads a few degrees high.
The analyzer measures the apparent contact angle in the image plane. For rough or patterned surfaces, measure several drops and viewing directions.
Reproducible results
The CSV export records every setting with each result: baseline method and coordinates, fit model, edge threshold, fit region, number of fitted points, image size and analyzer version. Anyone can repeat the measurement from the same image.
How it compares
The analyzer is meant for researchers who already have drop images and want a careful, documented measurement without installing software. Instrument software from the major manufacturers does much more, because it also controls the camera and the dosing, while the free ImageJ plugins are a manual route: the user places the contour and the baseline by eye on each image, which is slow and makes the result depend on the operator. The table summarises the main differences as described in each provider's public documentation.
| SOREC analyzer | Manual measurement in ImageJ (e.g. DropSnake, LB-ADSA) | Instrument software (e.g. KRÜSS ADVANCE, Biolin OneAttension, Ossila) | |
|---|---|---|---|
| Cost and access | Free, runs in any browser, nothing to install | Free, needs ImageJ on a computer | Supplied with the manufacturer's instrument |
| Images | Any side-view image (JPG, PNG) | Any image opened in ImageJ | Mainly images recorded with the instrument |
| Drop outline and baseline | Automatic, with manual baseline correction | Contour points and baseline placed by hand, so results depend on the operator and are prone to user error | Automatic, with manual correction |
| Fit models | Young–Laplace, local tangent | B-spline (DropSnake), low-Bond Young–Laplace (LB-ADSA) | Several, including Young–Laplace, conic section, polynomial and circle |
| Static angle, left and right | Yes | Yes | Yes |
| Advancing and receding angles, hysteresis, roll-off | As a SOREC analysis service | No dedicated workflow | Yes, with the instrument's dosing or tilting stage |
| Surface free energy | As a SOREC analysis service | No | Yes (an optional module on some systems) |
| Batch processing and CSV export | Yes | One image at a time, by hand | Yes |
| Open validation data | 80 published images, results in the validation note | Published method papers | Manufacturer's specifications |
KRÜSS, Biolin Scientific, Ossila and ImageJ are the names of their respective owners and are used here only to describe their products. Features may differ between instrument models and software versions.
How to cite
If you use the analyzer in published work, please cite it as:
Omeje, I. S. (2026). SOREC Contact Angle Analyzer (Version 2.0) [Computer software]. Surface Optimization Research & Consulting (SOREC) LLC, Hattiesburg, Mississippi, USA. https://soreclabs.com/contact-angle.html
BibTeX:
@software{omeje2026sorec,
author = {Omeje, Ilemona S.},
title = {{SOREC Contact Angle Analyzer}},
version = {2.0},
year = {2026},
publisher = {Surface Optimization Research \& Consulting (SOREC) LLC},
address = {Hattiesburg, Mississippi, USA},
url = {https://soreclabs.com/contact-angle.html}
}
Advanced wettability analysis
Advancing and receding contact angles, contact angle hysteresis, roll-off angles from tilted-plate images and surface free energy from several probe liquids are available from SOREC as an analysis service. Send us your images or videos and we return the results with a short report.
Ask about an analysisRelease history
| Version | Date | Changes |
|---|---|---|
| 2.0 | November 2026 | New analysis engine: background normalisation, sub-pixel edge at local half contrast, contact points found on the drop outline (substrate corner or reflection neck), Young–Laplace profile fit. Validated on 80 published experimental images. Fit model choice, inactive panel for advanced analyses, subscripted angle labels. |
| 1.0 | November 2026 | First release: threshold edge detection, baseline from substrate edge or reflection symmetry, polynomial tangent fit. Tested on synthetic images only. |
Results from different versions can differ. The CSV export records the version used for each measurement.
References
- J.-W. Lee, Paired sessile-drop contact angle measurements and analysis images from 80 water-droplet photographs, Mendeley Data, V1 (2026). doi:10.17632/t25zyc857h.1 (validation dataset)
- T. Huhtamäki, X. Tian, J.T. Korhonen, R.H.A. Ras, Surface-wetting characterization using contact-angle measurements, Nature Protocols 13 (2018) 1521–1538. doi:10.1038/s41596-018-0003-z
- A.F. Stalder, T. Melchior, M. Müller, D. Sage, T. Blu, M. Unser, Low-bond axisymmetric drop shape analysis for surface tension and contact angle measurements of sessile drops, Colloids and Surfaces A 364 (2010) 72–81. doi:10.1016/j.colsurfa.2010.04.040
- A.F. Stalder, G. Kulik, D. Sage, L. Barbieri, P. Hoffmann, A snake-based approach to accurate determination of both contact points and contact angles, Colloids and Surfaces A 286 (2006) 92–103. doi:10.1016/j.colsurfa.2006.03.008
- R. Akbari, C. Antonini, Contact angle measurements: from existing methods to an open-source tool, Advances in Colloid and Interface Science 294 (2021) 102470. doi:10.1016/j.cis.2021.102470
- J.M. Bland, D.G. Altman, Statistical methods for assessing agreement between two methods of clinical measurement, The Lancet 327 (1986) 307–310. doi:10.1016/S0140-6736(86)90837-8 (agreement statistics in the validation)
The Young–Laplace profile is integrated as dx/ds = cos φ, dz/ds = sin φ, dφ/ds = 2/b + cz − sin φ/x, with apex radius b and capillary constant c, and fitted to the drop outline. Details are in the validation note.